G3 (From G2) Improvements & Features
- Fully peer-reviewed system with certifications on electrical parts
- Single head, full-bore configuration for each system – promotes molecular flow, reduces operating temperature, and gives users increased ease-of-use compared to multi-head configurations
- Custom-cut ground glass drip tips – eliminate grease contamination and allow distillate to drip directly into the receiving flask
- All collection joints have been increased to 29/42 (29 mm) – to reduce potential clogging
- Condenser length has been extended for increased surface area and to accommodate larger collection flasks
- Improved heating mantle system with digital temperature and stir control
- Dual Temperature Monitoring allows the user to monitor temperature from two separate areas on the system such as the distillation head and cold trap, which improves thermal efficiency throughout the system, and allows for more producibility between runs
- Integrated supports built into each mantle – to make setup easy
- Full lattice rack included with each kit
- Reinforced mantle base – increases stability, dampens vibration, and provides better alignment for the whole system
- Improved glassware connections – utilizing ground glass joint to KF25 adapters for maximum vacuum capability and leak reduction
- Pinched thermocouple adapters for easy removal
- 1 year EliteLab® subscription included with each kit
G3X Improvements & Features
G3X kits include all of the improvements and features of the G3 kits as well as the following:
- Angled vacuum take-off valves – for both the swingarm and secondary collection port which allows for isolation from the rest of the system as well as unlimited fraction collection
- Secondary, liquid-cooled “cold finger” condenser – for greater condensation efficiency
- Jacketless primary condenser – for increased flow; keeps volatiles in vapor state for collection away from the main body fraction
Complete System
Peer-Reviewed, GMP Compliant
Heating Mantles
CSA
CAN/CSA C284101, C22.2 No. 88
Temperature Controller
NRTL, QPS, CE, Constructed with 100% UL and CSA listed parts.
UL 61010-1; 61010-2-201
CAN/CSA 61010-1, C22.2
Temperature Monitor
CE, Constructed with 100% UL and CSA listed parts.
Stir Controller
CE, Constructed with 100% UL and CSA listed parts.
Vacuum Monitor
RoHS / CE / NRTL
Glassware
Meets ASTM specification E438, type 1, class A, Kimble Kimax or Schott Duran Glass
G3/G3X Specs: | 5L | 12L | 20L | 20L G3X-R |
Batch Capacity: | 2500mL (approx. 2500g dewaxed oil) | 6000mL (approx. 6000g dewaxed oil) | 10,000mL (approx. 10,000 grams dewaxed oil) | 10,000mL (approx. 10,000g dewaxed oil) |
Optimum Capacity: | 900 – 1500mL | 3000 – 4000mL | 7,000 – 9,000mL | 7,000 – 9,000mL |
Absolute Max. Capacity: | 3000mL | 8000mL | 12,0000mL | 12,000mL |
Main Body Fraction Flow Rate: | 400 – 800 grams/hour | 1000 – 2000 grams/hour | 2000 – 4000 grams/hour | 2,000 – 4,500 grams/hour |
Batch Duration: | 4 – 8 Hours
(estimated) |
4 – 10 Hours
(estimated) |
4 – 10 Hours
(estimated) |
4-10 Hours
(estimated) |
Benchtop Size: | 30″ x 30″ x 30″ (not including pump/trap/circulator) | 43″ x 27″ x 40″ (not including pump/trap/circulator) | 45″ x 27″ x 42″ (not including pump/trap/circulator) | 45″ x 27″ x 42″ (not including pump/trap/circulator) |
Distillation Head: | PDH-2S (silvered) | PDH-3 (silvered) | PDH-6 (silvered) | DH-150HE |
G3/G3X Requirements: | 5L | 12L | 20L | 20L G3X-R |
Vacuum Pump: | Pair with LS-RV-12, LS-RV-18, LS-RV-25, or LS-RV-36, CRVpro16, CRVpro24, CRVpro30 | Pair with LS-RV-12, LS-RV-18, LS-RV-25, or LS-RV-36, CRVpro16, CRVpro24, CRVpro30 | Pair with LS-RV-12, LS-RV-18, LS-RV-25, or LS-RV-36, CRVpro16, CRVpro24, CRVpro30 | Pair with: LS-RV-18, LS-RV-25, LS-RV-36, LS-RV-50, CRVpro16, CRVpro24, CRVpro30 |
Cold Trap: | Glass: LS-DCT-2HE
Mechanical: MAXTRAP 105 |
Glass: LS-DCT-2HE
Mechanical: MAXTRAP 105 |
Glass: LS-DCT-2HE
Mechanical: MAXTRAP 105 |
Glass: N/A
Mechanical: MAXTRAP 105 |
Liquid Circulator: | MX07R, SD07R, AD07R, K-6, CD-200F | MX07R, SD07R, AD07R, K-6, CD-200F | MX07R, SD07R, AD07R, K-6, CD-200F | MX07R, SD07R, AD07R, Kiss K-6, CD-200F
Chiller Temp. Requirement: 32°F, 0°C or below |
G3/G3X Electrical Parameters: | 5L | 12L | 20L | 20L G3X-R |
Heating Mantle: | 1150w, 120V, 10 amps/ EU 950w, 230V, 5 amps
Max temp: 260°C (PTFE/PFA temp. limit) |
1750w, 120V, 15.2 amps / EU 1750w, 230V, 7.7 amps
Max temp: 260°C (PTFE/PFA temp. limit) |
1750w, 120V, 15.2 amps / EU 1750w, 230V, 7.7 amps
Max temp: 260°C (PTFE/PFA temp. limit) |
1750w, 120V, 15.2 amps / EU 1750w, 230V, 7.7 amp
Max temp: 260°C (PTFE/PFA temp. limit) |
Stirring Motor: | 110 – 240V, 50/60 Hz, 2 amps. Speed Range: 100 – 1850 RPM | 110 – 240V, 50/60 Hz, 2 amps. Speed Range: 100 – 1850 RPM | 110 – 240V, 50/60 Hz, 2 amps. Speed Range: 100 – 1850 RPM | 110 – 240V, 50/60 Hz, 2 amps. Speed Range: 100 – 1850 RPM |
Vacuum Monitoring: | 100 – 240V, 50/60 Hz, 0.3 amps
Range: 1 mTorr – 760 Torr |
100 – 240V, 50/60 Hz, 0.3 amps
Range: 1 mTorr – 760 Torr |
100 – 240V, 50/60 Hz, 0.3 amps
Range: 1 mTorr – 760 Torr |
100 – 240V, 50/60 Hz, 0.3 amp
Range: 1 mTorr – 760 Torr |
Temperature Monitoring: | 100 – 240V, 50/60 Hz, 1 amp | 100 – 240V, 50/60 Hz, 1 amp | 100 – 240V, 50/60 Hz, 1 amp | 100-240V, 50/60 Hz, 1 amp |
Precision offers the following on-site product installation and training with experienced extraction technicians:
- Extract Product Engineering
- Distillation Prep
- Distillation Training