Systems Include:
- COLDWELL 304 Stainless Steel Trap – patent pending baffle system.
- 3 Quart Capacity (2839 mliters)
- Viton Gasket
- Labconco -50°C / -105°C Refrigerated Base
- Green CFC / HFC Free Refrigerants
- KF-25 Inlet/Outlet, 1 each 4′ Connection Line with KF-25 flanges, clamps, centering ring
- Power Cord
- 3mm Neoprene Insulator Cap
- 1 Year Warranty
*Requires 800ml (~ 3.5 cups) of a transfer fluid such as ethanol. Not Included. recommends: ethanol, glycol, cryo-cool fluids
MAXTRAP50 COLD TRAP
DIMENSIONS
13.5″ W x 23.5″ D x 12.0″ H
WEIGHT
COLDWELL Stainless Trap: 14 lbs
Thermo Mechanical Base: 60 lbs
Shipping Weight: 95 lbs
MAXTRAP105 COLD TRAP
DIMENSIONS
24.1″ W x 23.9″ D x 12.6″ H
WEIGHT
COLDWELL Stainless Trap: 14 lbs
Mechanical Base: 139 lbs
Shipping Weight: 190 lbs
COLDWELL STAINLESS TRAP
KF25 Inlet & Outlet
3 Quart Capacity (2839 mliters)
Viton Gasket
Neoprene Insulator Cap
LABCONCO -50°C / -105°C MECHANICAL BASE
115V / 1ph / 3 AMP (60Hz)
Cord Included
CFC / HCFC Free Refrigerants
CONNECTIONS
1 each 4′ KF-25 flanged vacuum rated line
1 each KF-25 Centering Ring
1 each KF-25 Clamp
- 1 Year Warranty
MAXTRAP Tips:
Safety
At very cold temperatures, solvents and even oxygen can re-condense back into liquid. This is potentially explosive. Liquid oxygen is a light blue color and can condense into a cold trap if a pump has sucked air through the trap when the trap is very cold. Always follow your standard operating procedures and safety guidelines.
Don’t Stop: Keep Processing while trap is being reclaimed, cleaned, refreshed
Cold traps can become full or the captured material freezes and blocks flow. For high moisture, outgassing, 24 hour processing/vacuum purging consider adding a second, spare COLDWELL trap.
The secondary COLDWELL can be dropped into place so processing continues while the full trap is cleaned, defrosted.
Quick Start Up Tips:
- Remove shipping foam from reservoir bottom prior to use
- Fill reservoir with 800ml (+/-3.5 cups) ethanol to marked line
- Monitor ethanol levels inside reservoir
- Allow at least 30 mins for trap to cool
- Keep side vents clear. Allow at least 4″ for airflow
- Monitor interior of stainless trap for clogging
- Short tube – attach to vacuum pump
- Tall tube – attach to vacuum oven